CHLORINE, TOTAL

Chlorine, Total

10 mL

SPECIFICATIONS

Range

0.00 to 5.00 mg/L Cl

2

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L or ±4 % of reading @ 25 °C, whichever is greater

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the EPA method 330.5 and

Standard Methods for the Examination of

Water and Wastewater, 20

th

edition, 4500-Cl G, DPD method. The reaction between

chlorine and the DPD reagent causes a pink tint in the sample.

REQUIRED REAGENTS

Code

Description

Q.ty/test

HI 93711-0

DPD Powder Reagent

1 packet

REAGENT SETS

HI 93711-01

Reagents for 100 tests

HI 93711-03

Reagents for 300 tests

For other accessories see page 68.

MEASUREMENT PROCEDURE

• Select

Chlorine, Total method following one of the procedures described in the “Method Selection”

section (see page 15).

• Take an empty vial.

• Fill the vial with 10 mL of unreacted sample, then replace the cap. This is the blank.

• Place the vial into the holder and push it completely down.

• Press the Zero key. The instrument will perform a zero sequence and after a few seconds the display

will show “

-0.0-”. Now the meter is zeroed and ready for measurement.