Hanna Instruments HI 83214 User Manual

Page 39

Advertising
background image

39

• Press Start and the display will show the countdown prior to the measurement and the ”Reaction

Time” message. Alternatively, wait for 5 minutes and press Zero.

• The instrument will perform a zero sequence and after a few seconds the display will show

-0.0-”. Now the meter is zeroed and ready for measurement.

• Remove the blank vial and place the sample vial into the holder and push it completely down.
• Press Read. The instrument will perform the reading. The instrument displays the concentration in

mg/L of total nitrogen (N).

• Press ▲ or ▼ to convert the result in mg/L of ammonia (NH

3

) or mg/L of nitrate (NO

3

-).

The method detects all organic and inorganic forms of nitrogen present in the sample.

INTERFERENCES

Interference may be caused by:
Bromide (Br-) above 240 mg/L
Chloride (Cl-) above 3000 mg/L
Chromium (Cr

3+

) above 0.5 mg/L

Nitrogen, Total HR