Hanna Instruments HI 83214 User Manual

Page 59

Advertising
background image

59

PHOSPHORUS, REACTIVE HIGH RANGE

Phosphorus, Reactive HR

SPECIFICATIONS

Range

0.0 to 32.6 mg/L P

Resolution

0.1 mg/L

Accuracy

±0.5 mg/L or ±5 % of reading @ 25 °C, whichever is greater

Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

20

th

edition, 4500-P C, vanadomolybdophosphoric acid method. The reaction between

orthophosphate and the reagents causes a yellow tint in the sample.

REQUIRED REAGENTS

Code

Description

Q.ty/test

Q.ty/set

HI 93763A-0*

Reagent Vial

1 vial

50 vials

DEIONIZED120

Deionized Water

5 mL

1 bottle

*

Reagent Vial identification: P RHR, green label.

Note: Store the unused vials in their container in a cool and dark place.

REAGENT SET

HI 93763A-50 Reagents for up to 49 tests
For other accessories see page 68.

MEASUREMENT PROCEDURE

Reagent Blank Correction: This method requires a reagent blank correction. A single blank vial may be used
more than once; the blank vial is stable up to two weeks (room temperature). For most accurate measurement,
run a blank for each set of measurements and always use the same lot of reagents for blank and samples.

• Select the

Phosphorus, Reactive High Range method following one of the procedures described in the

“Method Selection” section (see page 15).

• Remove the caps from two Reagent Vials.

• Add exactly 5.0 mL of sample to one vial (sample vial), and 5.0 mL of deionized water to the other

vial (blank vial), while keeping the vials at a 45-degree angle.

green label