34

• Press Start and the display will show the countdown prior to the measurement and the “Reaction

Time” message. Alternatively, wait for 5 minutes and press Zero.

• The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the blank vial.

• Place the sample vial into the holder and push it completely down.

• Press

Read. The instrument will perform the reading. The instrument displays concentration in mg/L of

total nitrogen (N).

• Press ▲ or ▼ to convert the result in mg/L of ammonia (NH

3

) or mg/L of nitrate (NO

3

-).

The method detects all organic and inorganic forms of nitrogen present in the sample.

INTERFERENCES

Interference may be caused by:
Bromide (Br-) above 60 mg/L
Chloride (Cl-) above 1000 mg/L
Chromium (Cr

3+

) above 0.5 mg/L

Nitrogen, Total LR