Green label – Hanna Instruments HI 83214 User Manual

Page 47

Advertising
background image

47

• Choose a homogeneous sample. Samples containing settleable solids need to be homogenized with a

blender.

• Preheat the Hanna Reactor HI 839800 to 150 °C (302 °F). For correct use of the reactor follow Reactor

Instruction Manual.
Use of the optional HI 740217 safety shield is strongly recommended.
DO NOT USE AN OVEN OR MICROWAVE samples may leak and generate a corrosive and possibly explosive
atmosphere.

• Remove the caps from two Reagent Vials.

• Add exactly 0.2 mL of sample to one vial (sample vial), and 0.2 mL of deionized water to the other

vial (blank vial), while keeping the vials at a 45-degree angle. Replace the cap tightly and mix by
inverting each vial a couple of times.

Warning: The vials will become very hot during mixing, be careful

when handling them.

• Insert the vials into the reactor and heat them for 2 hours at 150 °C.

• At the end of the digestion period switch off the reactor. Wait for twenty

minutes to allow the vials to cool to about 120 °C.

• Invert each vial several times while still warm, then place them in the

test tube rack.

Warning: The vials are still hot, be careful when handling them.

• Leave the vials in the tube rack to cool to room temperature. Do not shake

or invert them anymore otherwise the samples may become turbid.

Blank

Sample

Sample

Blank

Blank

D.I. Water

Sample

Sample

COD HR

green label