Nitrogen, Total HR

Sample

• Replace the cap tightly and shake vigorously the vials for about 30 seconds until all the powder is

completely dissolved.

• Insert the vials into the reactor and heat them for 30 minutes at 105 °C.

Note: To obtain most accurate results, it is strongly recommended to

remove the vials from the reactor after 30 minutes.

• At the end of the digestion place the vials in the test tube rack and allow to

cool to room temperature.
Warning: As the vials are still hot, be careful in handling them.

• Select

Nitrogen, Total HR method following one of the procedures described in the “Method Selection”

section (see page 15).

• For this method the instrument provides 3 reaction timers which can be used throughout the procedure.

• Remove the cap from the vials and add the content of one packet of Sodium Metabisulfite for Total

Nitrogen analysis to each vial. Replace the cap tightly and shake gently the vials for 15 seconds.

• Press Timer to start the 3 minutes timer or wait for 3 minutes (without shaking the vials) to allow the

reaction to complete.