NITROGEN, TOTAL HIGH RANGE

Nitrogen, Total HR

SPECIFICATIONS

Range

10 to 150 mg/L N

Resolution

1 mg/L

Accuracy

±3 mg/L or ±4 % of reading @ 25 °C, whichever is greater

Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Chromotropic acid method. A persulfate digestion converts all forms of nitrogen to
nitrate. Then the reaction between nitrate and the reagents causes a yellow tint in
the sample.

REQUIRED REAGENTS

Code

Description

Q.ty/test

Q.ty/set

HI 93767B-B*

Digestion Vial

1 vial

50 vials

DEIONIZED120

Deionized Water

0.5 mL

1 bottle

PERSULFATE/N

Potassium Persulfate

1 packet

50 packets

BISULFITE/N

Sodium Metabisulfite

1 packet

50 packets

HI 93767-0

Total Nitrogen Reagent

1 packet

50 packets

HI 93766V-0HR**

Reagent Vial

1 vial

50 vials

*

Digestion Vial identification: N HR, red label.

**

Reagent Vial identification: N HR, green label.

Note: Store the unused vials in their container in a cool and dark place.

REAGENT SET

HI 93767B-50 Reagents for up to 49 tests. Contains:

Box 1: HI 93767B-50 Reagent Set
Box 2: HI 93767A&B-50 Reagent Set, for Total High Range Nitrogen.

For other accessories see page 68.

REQUIRED ACCESSORIES

HI 839800-01 Hanna reactor (115 VAC)
HI 839800-02 Hanna reactor (230 VAC)
HI 740216

Test tube cooling rack (25 holes)

HI 740217

Laboratory bench safety shield

For other accessories see page 68.