Hanna Instruments HI 83214 User Manual

Page 29

Advertising
background image

29

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the vial.
• Remove the cap and add the content of one packet of HI 93766-0 Nitrate

Reagent.

• Replace the cap tightly and invert the vial 10 times. This is the reacted

sample.
Note: The method is technique sensitive: to obtain most reproducible results

it is strongly recommended to follow carefully the “invert” procedure
described on page 10.

• Place the vial into the holder and push it completely down.
• Press

Timer and the display will show the countdown prior to the measurement and the ”Reaction

Time” message. Alternatively, wait for 5 minutes and press Read. The instrument will perform the
reading.

• The instrument displays concentration in

mg/L of nitrate-nitrogen (NO

3

--N). Press ▲ or ▼ to

convert the result in mg/L of nitrate (NO

3

-).

INTERFERENCES

Interference may be caused by:
Barium (Ba

2+

) above 1 mg/L

Chloride (Cl-) above 1000 mg/L
Nitrite (NO

2

-) above 50 mg/L

Samples containing up to 100 mg/L nitrite may be measured after the following treatment: add 400 mg
of urea to 10 mL of sample, mix until completely dissolved, then proceed with the usual measurement
procedure.

Nitrate