PHOSPHORUS, ACID HYDROLYZABLE

Phosphorus, Acid Hydrolyzable

SPECIFICATIONS

Range

0.00 to 1.60 mg/L P

Resolution

0.01 mg/L

Accuracy

±0.05 mg/L or ±5 % of reading @ 25 °C, whichever is greater

Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the EPA method 365.2 and

Standard Methods for the Examination of

Water and Wastewater, 20

th

edition, 4500-P E, ascorbic acid method. A mild acid

digestion converts condensed inorganic forms of phosphates to orthophosphate. Then
the reaction between orthophosphate and the reagents causes a blue tint in the
sample.

REQUIRED REAGENTS

Code

Description

Q.ty/test

Q.ty/set

HI 93758V-0AH*

Reagent Vial

1 vial

50 vials

HI 93758B-0

NaOH Solution 1.20 N

2 mL

1 bottle

HI 93758-0

Phosphorus Reagent

1 packet

50 packets

*

Reagent Vial identification: P AH, white label.

Note: Store the unused vials in their container in a cool and dark place.

REAGENT SET

HI 93758B-50 Reagents for 50 tests
For other accessories see page 68.

REQUIRED ACCESSORIES

HI 839800-01 Hanna reactor (115 VAC)
HI 839800-02 Hanna reactor (230 VAC)
HI 740216

test tube cooling rack (25 holes)

HI 740217

Laboratory bench safety shield

For other accessories see page 68.

MEASUREMENT PROCEDURE

Before using the reagent kit carefully read all the instructions and the Material Safety Data Sheet

(MSDS). Pay particular attention to all warnings, cautions and notes. Failure to do so may result

in serious injury to the operator.

• Preheat the Hanna Reactor HI 839800 to 150 °C (302 °F). For correct use of the reactor follow Reactor

Instruction Manual.
The optional HI 740217 safety shield is strongly recommended.
DO NOT USE AN OVEN OR MICROWAVE samples may leak and generate a corrosive and possibly explosive
atmosphere.