Hanna Instruments HI 83214 User Manual

Page 64

Advertising
background image

64

Blank

Sample

Blank

Sample

Blank

Sample

Blank

Sample

Phosphorus, Total HR

• Select the

Phosphorus, TTTTTotal High Range method following one of the procedures described in the

“Method Selection” section (see page 15).

• Remove the cap from the vials and add exactly 2.0 mL of Sodium Hydroxide (NaOH) Solution 1,54 N

to each vial, while keeping the vials at a 45-degree angle. Replace the cap tightly and mix by
inverting the vials a couple of times.

• Remove the cap from the vials and add exactly 0.5 mL of HI 93763B-0 Molybdovanadate Reagent

to each vial, while keeping the vial at a 45-degree angle. Replace the cap tightly and mix by
inverting the vials a couple of times.

• Place the blank vial into the holder and push it completely down.

• Press Timer and the display will show the countdown prior to the measurement and the “Reaction

Time” message. Alternatively, wait for 7 minutes and press Zero.

• The instrument will perform a zero sequence and after a few seconds the display will show “-0.0-”. Now

the meter is zeroed and ready for measurement.

Advertising