Hanna Instruments HI 83214 User Manual

Page 42

Advertising
background image

42

• Select

Oxygen Demand, (COD) LR method following one of the procedures described in the “Method

Selection” section (see page 15).

• Place the blank vial into the holder and push it completely down.

• Press the Zero key. The instrument will perform a zero sequence and after a few seconds the display

will show “

-0.0-”. Now the meter is zeroed and ready for measurement.

• Remove the blank vial.

• Place the sample vial into the holder and push it completely down.

• Press

Read. The instrument will perform the reading.

• The instrument displays concentration in

mg/L of oxygen demand.

INTERFERENCES

Interference may be caused by:
Chloride (Cl-) above 2000 mg/L.
Samples with higher chloride concentration should be diluted.

COD LR

Advertising