Chapter 3: Getting Started

Product Service Domestic 1-800-522-7658, International 816-333-8811

Never lift the lid if the block is
moving. Samples may splash out.

Never operate the RapidVap without
the sample block securely fastened in
place. Excessive vibration will occur.

Never attempt to reach into the
chamber if the block is moving.























Nitrogen Gas Connection


Turn the gas supply off. Install a barb fitting suitable
for a 1/4" I.D. hose into the “out” port of the regulator.
Clamp securely. Alternately, a compression fitting may
be used that is suitable for a hose with a 3/8" O.D. and
1/4" I.D.

THE NITROGEN PRESSURE MUST
NEVER BE ALLOWED TO EXCEED
20 PSI. (103 KPA)

!

N

)

Nut or guide rods – 3 required

Washer – 3 required

Block

Mounting stud

Temperature sensor

Apply heat
transfer grease