Determination of Operating Conditions
Applied Vacuum/Boiling Point:
t
s
= Boiling point at 1 bar (normal pressure)
t
p
= Boiling point at P mbar (operating pressure)
p = Operating pressure in mbar
Example methyl alcohol:
1. Boiling point at normal (1 bar) atmospheric
pressure = 65°C.
2. Constant b = 0.160
3. Operating pressure = 130 mBar
4. Boiling point at 130 mBar pressure = 24°C
Advertising
This manual is related to the following products:
- RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910014 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910000 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900003 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900002 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900000