Chapter 7: Troubleshooting

Product Service Domestic 1-800-522-7658, International 816-333-8811


PROBLEM

CAUSE

CORRECTIVE ACTION

Sample in
bottom of
chamber

Gas pressure too high
(RapidVap N

2

)

Reduce pressure.

Vortex speed too high

Reduce speed.

Condensation forming
inside lid (RapidVap
Vacuum)

Install lid heater. See page 36.

Poor recovery
rates

Sample going to
dryness (RapidVap N

2

)

Program RapidVap N

2

to prevent dryness.

Sample sloshing out of
tube

Decrease vortex speed.

Heat setting is too high Decrease heat.

Evaporation
rate is reduced

Gas supply depleted
(RapidVap N

2

)

Replenish gas.

Heater inoperable

Contact Labconco.

Vacuum pump failure
(RapidVap Vacuum)

Check pump.

No vacuum
poor vacuum
(RapidVap
Vacuum)

Pump not on

Turn on pump.

Control valve open

Check control valve.

Leaks in lines or
connectors

Locate and repair.

Foreign material on lid
gasket

Clean gasket and lid.

Vacuum pump oil low

Add oil.

Excessive
contamination in
vacuum pump oil

Replace oil.

Pump not connected to
RapidVap

Connect hose and power cord.