50

Chapter 7: Troubleshooting

Product Service Domestic 1-800-522-7658, International 816-333-8811

DISPLAY
ERROR CODE

CAUSE CORRECTIVE ACTION

Mem Fail – Push
Prog

Memory failure

Push program button.

New memory IC
chip

Push program button.

Bad memory IC
chip

Call Labconco – Replace IC chip or control
PCB.

Motor Error

Stalled condition

Turn power off. Try to restart after one
minute. Call Labconco if error message
remains.

Excessive speed
detected, i.e. >120%

Turn power off. Try to restart after one
minute. Call Labconco if error message
remains.

Defective motor

Turn power off. Try to restart after one
minute. Call Labconco if error message
remains.



Other corrective actions for potential problems are as
follows:

PROBLEM

CAUSE

CORRECTIVE ACTION

Unit will not
operate

Unit not connected to
electrical power

Connect unit to proper electrical
receptacle.

Circuit breaker blown

Correct electrical problem and reset
circuit breaker by pressing button.

Lid open

Close lid.

Excessive
vibration

Sample tubes not
located symmetrically
in block

Reposition sample tubes.

Sample block loose

Tighten all three mounting screws.

Sample block not
installed

Install sample block.

Sample odor in
lab

Vent hose exhausting
into lab area

Redirect hose to fume hood.

Inoperable blower
(RapidVap N

2

)

Replace blower.