Refer to the following if your RapidVap fails to operate
properly. If the suggested corrective actions do not
solve your problem, contact Labconco for additional
assistance. The following failure codes may appear on
the display when problems are sensed by the internal
self check routine.

DISPLAY
ERROR CODE

CAUSE CORRECTIVE ACTION

Top Temp Sensor

Top sensor failure

Replace top sensor assembly.

Connection failure Repair connection.

Bottom Temp
Sensor

Bottom sensor
failure

Replace bottom sensor assembly.

Connection failure Repair connection.

Vacuum Sensor

Sensor failure

Replace sensor assembly.

Connection failure Repair connection.

Lid Opened

Lid open

Close lid.

Switch failure

Replace switch.

Connection failure Repair connection.