Appendix C: RapidVap Specifications

Product Service Domestic 1-800-522-7658, International 816-333-8811


evaporate from the top of the stem to 0.5 ml
is 17 minutes. Adding the two times
together, the total time to evaporate 300 ml
of methylene chloride to 0.5 ml in a sample
tube with a 1.5 ml stem is approximately 76
minutes.

If fewer than eight samples are run at a time,
the time to evaporate will become shorter.








Solvent:

Methylene

Chloride

Ethyl

Acetate

Hexanes

Acetone

Toluene

Water

Block Preheat &
Block Set Point
Temperature:


40°C


75°C


70°C


55°C


100°C


95°C

Starting Volume

450 ml

91

82

54

90

109

623

400

80 65 52 75 80 531

350

69 56 35 64 69 437

300

59 48 29 54 60 379

250

51 40 22 45 50 323

200

43 33 19 38 41 270

150

36 26 16 31 32 221

100

28 19 11 23 23 173

50

18 12 7 14 13 106











Note 1:
Even though the RapidVap N

2

is capable of

processing starting volumes as high as 450 ml using the
600 ml sample vial, its optimum volume is 300 ml. Above

Table A

Evaporation Time in Upper Portion of Sample Vial

(600 ml Sample Tubes) (Time in Minutes)

N

Number of samples: 8
Nitrogen pressure: 10 psi at volumes 0 thru 300 ml
(see Note 1)

5 psi at volumes above 300 thru 350 ml

4 psi at volumes above 350 thru 400 ml

2 psi at volumes above 400 thru 450 ml

Vortex speed: 70% at volumes 0 thru 300 ml
(see Note 1) 36% at volumes above 300 thru 450 ml
Sample tubes covered with caps