This Appendix contains technical information about the
RapidVap including specifications, environmental
operations conditions, wiring diagrams and evaporation
rates.
Electrical Specifications
• Nominal amperage for 115V RapidVap Vacuum
(model 79000-00 & 79000-10) excluding vacuum
pumps: 8A
Maximum amperage with vacuum pump: 16A
• Nominal amperage for 230V RapidVap Vacuum
(model 79000-01 & 79000-11) excluding vacuum
pumps: 4A
Maximum amperage with vacuum pump: 8A
• Nominal amperage for 115V RapidVap Vacuum
(model 79000-02 & 79000-12) with Lid Heater
excluding vacuum pump: 8.6A
Maximum amperage with vacuum pump: 16A
• Nominal amperage for 230V RapidVap Vacuum
(model 79000-03 & 79000-13) with Lid Heater
excluding vacuum pump: 4.3A
Maximum amperage with vacuum pump: 8A
• Nominal amperage for 115V RapidVap N
2
(model
79100-00, 79100-10, 79100-12 & 79100-14): 8A
• Nominal amperage for 230V RapidVap N
2
(model
79100-01, 79100-11, 79100-13 & 79100-15): 4A
• Frequency (all models): 50/60 Hz
• Phase: Single
• Heater power: 1000 watts