This Appendix contains technical information about the
RapidVap including specifications, environmental
operations conditions, wiring diagrams and evaporation
rates.

Electrical Specifications

• Nominal amperage for 115V RapidVap Vacuum

(model 79000-00 & 79000-10) excluding vacuum
pumps: 8A

Maximum amperage with vacuum pump: 16A

• Nominal amperage for 230V RapidVap Vacuum

(model 79000-01 & 79000-11) excluding vacuum
pumps: 4A

Maximum amperage with vacuum pump: 8A

• Nominal amperage for 115V RapidVap Vacuum

(model 79000-02 & 79000-12) with Lid Heater
excluding vacuum pump: 8.6A

Maximum amperage with vacuum pump: 16A

• Nominal amperage for 230V RapidVap Vacuum

(model 79000-03 & 79000-13) with Lid Heater
excluding vacuum pump: 4.3A

Maximum amperage with vacuum pump: 8A

• Nominal amperage for 115V RapidVap N

2

(model

79100-00, 79100-10, 79100-12 & 79100-14): 8A

• Nominal amperage for 230V RapidVap N

2

(model

79100-01, 79100-11, 79100-13 & 79100-15): 4A

• Frequency (all models): 50/60 Hz

• Phase: Single

• Heater power: 1000 watts

• Lid heater power: 40 watts