The log function
To be able to print a report the CPR data first need to be
logged by activating the log function. The log function can
be activated either on the SkillReporter if connected to the
manikin, or on the manikin. To activate the log function on
the SkillReporter you first have to connect it to the manikin
(the SkillReporter has no internal power source), then press the
“On/Off” button once and after the selftest is complete press
the “LOG” C button.
If you decide to use the manikin without the SkillReporter
attached, press the “On/Off” button at the right side of the
torso and then press the “LOG”

C button to activate the log

function.
To stop the log function press the “LOG” button a second
time. The CPR data is now logged and can be printed multiple
times. Just connect the SkillReporter and press one of the print
buttons. The data will not be lost even if the SkillReporter and
the manikin are switched off. When starting a new log, the
current log will be cleared and
data cannot be retrieved.

Printing a report

Short report D
The short report format
(statistical data), can only be
printed after the CPR(-D) data
has been logged. If the “short
print” button is pressed during
logging of a scenario, a sum-
mary report (statistical data) of
the performance will be printed
with data collected up to the
time of activating the short
report. If the “short print” but-
ton is pressed after the logging
of a scenario has been com-
pleted, a summary report of the
performance will be printed
with data for the whole session.

SkillReporter

Consistent and objective
real time feedback on
performance during practice
helps develop
and reinforce skills. The SkillReporter has a metronome B built
in providing a rate of 80 or 100 strokes/minute, and provides
feedback to the instructor and student to support the student’s
initial skill development by means of colour LEDs. In addition
the SkillReporter gives you the possibility to print a report of the
performance (either complete report with ventilation and
compression curves and a statistical report or a statistical
report only).

Note: In accordance with Guidelines 2010, use setting 100
compressions/minute.

To use
Turn the SkillReporter on by pressing the “on/off” A key either
on the SkillReporter or on the right side of the manikin torso.
Caution: Make sure the chest skin is properly fastened before
turning on the SkillReporter/manikin.
Pressing a second time will
turn the SkillReporter off. The SkillReporter and manikin will turn
off automatically if no activity takes
place for 10 minutes
(60 minutes if the log function is
active). Note: The manikin can
also be used without the
SkillReporter connected. You are
then able to log the training
scenario and later on reconnect
the SkillReporter to print a report.
The light indicators as described below will give instant feedback
on CPR performance.

A C

LED(s)

Indication

Limits

*


Inflation volume:

yellow

insufficient

<0.5 l

green

sufficient

0.5 l - 0.8 l

red

excessive

>0.8 l

Compression depth:

yellow

insufficient

<50 mm (2”)

green

sufficient

50 mm

Note: The maximum compression depth indicator
is de-activated on this product in accordance with
2010 Guidelines.


Too fast inflation:

red (sufficient ventilation < 1 sec. )

Wrong hand position:

yellow

Pressure outside

of acceptable area

(sides and

upper part)

red

Pressure outside

of acceptable area

(lower center)

Pulse

SkillReporter

ResusciAnne

*Measurement tolerance in the manikin: +/- 15% in normal room temperature.
Measurements based on ILCOR guidelines 2010.

Student name:

_______________________

Instructor name:

_______________________

Duration of the session: (min : sec)

____

Ventilations:
Average volume (ml)

____

Average number per min. (#/min)

____

Minute volume (ml/min)

____

Total number (#)

____

Number correct (#)

____

Percent correct (%)

____

Too much (#)

____

Too little (#)

____

Too fast (#)

____

Compr. / Vent. ratio

____

Compressions
Average depth (mm)

____

Average number per min. (#/min)

____

Average compression rate (#/min)

____

Total number (#)

____

Number correct (#)

____

Percent correct (%)

____

Too deep (#)

____

Too shallow (#)

____

Wrong hand position (#)

____

Hand position too low

(#)

____

Incomplete release (#)

____

Defibrillation (CPR-D models only)
Time from scenario start
to first shock (min : sec)

____ ____

Time from call for help
to first shock (min : sec)

____ ____

Time from arrival of defibrillator
to first shock (min : sec)

____ ____

Light indicators on the SkillReporter