6 Fit the hooks on the bottom of the clear

hard protective lid into the holes at
the bottom edge of the cartridge, and
press the lid and handle down (G).
Be sure the cartridge handle snaps
into place.The training cartridge is
now ready for its next use.

7 Reinstall the cartridge if necessary and press down the protective lid.

The Trainer will announce the cartridge type and turn on for scenario
selection. Press and hold the On/Off button or do nothing and the
Trainer will automatically go to the training standby mode.

Maximizing Training Pads Cartridge Service Life

The M5073A and M5074A Training Pads are designed for up to 100 uses.
To help ensure continued performance, avoid placing the resealable film
cover or adhesive side of the pads in heat, direct sunlight, or against surfaces
that might transfer fiber, dirt, or lint. Periodically check the metal contacts on
the back of the cartridge and clean them with a soft, non-abrasive cloth if
necessary. Do not use alcohol to clean these contacts. Do not clean the
contacts in the Trainer’s cartridge recess; they are self-cleaning.

When the original training pads have reached the end of their service life,
you can order M5093A Replacement Adult Training Pads or M5094A
Replacement Infant/Child
Training Pads for the cartridge.
These include a set of pads and
a resealable film cover. Change
the pads as shown in the figure.