Manikin Setup

ATTACHING THE LEFT ARM
Open the torso as described in steps 1 - 4, Opening the Torso. Follow the procedures listed below in reverse to detach the arms.

1 Align the left arm axle with the shoulder socket.

2 Ensure that the shoulder screw is loose enough to allow the arm axle to slide easily into place.

3 Feed the cables from the arm axle through the shoulder socket.

4 Carefully push the arm axle into the shoulder bracket, so that the axle is flush with the inside of the bracket.

LA PULSES

BP

LA

CONV

P

B

L

BP

PULSES

L

L

LA

B

5 Tighten the shoulder screw with the Allen wrench.

6 Connect the arm cables to the corresponding connection points in the torso.

LEFT ARM CONNECTIONS

Name/Label

Tube/Cable Color

Connector Description

LA Pulses

Grey cable

Black rectangular connector, 6 lead

Blood Pressure

Grey cable

Black rectangular connector, 2 lead

ATTACHING THE RIGHT ARM

Open the torso as described in steps 1 - 4, Opening the Torso.
Follow the procedures listed below in reverse to detach the arms.

1 Align the arm axle with the shoulder socket and feed the cables

from the arm axle through the shoulder socket

2 Push the axle into the bracket until it is flush with the inside of the

bracket.

3 Tighten the shoulder screw with the Allen wrench.

4 Connect the corresponding cables as shown below.

RIGHT ARM CONNECTIONS

Name/Label

Tube/Cable Color

Connector Description

Right Arm

Black harness cable

Black rectangular connector, 4 lead

Air

Transparent tube

White tube connector twist

Right Radial

Grey harness cable

Black with silver colored connector