Troubleshooting – Laerdal SimMan Essential User Manual

Page 68

Advertising
background image

68

Troubleshooting

Troubleshooting

SYSTEM SETUP

Problem

• Lost data or total system failure (General System failure)

Solution

• Should system shutdown or all data is lost or corrupted;

please contact your local Laerdal Service Center.

WLAN STABILITY AND PERFORMANCE:

Problem

Connection between the PCs and the manikin is frequently lost.

Causes

• The WLAN signal might interfere with other radio traffic or

noise.

• E.g. WLAN and Bluetooth networks, wireless telephones,

microwaves etc.

Possible Solutions

• Change WLAN channel see section: Changing WLAN

settings.

• Place manikin closer to the PCs. Avoid solid barriers which

may block the signal between the manikin and PCs.

• Windows includes power management features. Configure

the WLAN for Maximum performance. Some WLAN
adapter drivers come with their own power management
settings.

• When the manikin is stationary, consider using a LAN

network between manikin and PCs.

• Change to external router.

• Use Laerdal Connection Manager to locate the best channel.

WLAN CONNECTION

Problem

• The manikin works via LAN cable, but not with WLAN

connection.

Possible Solutions

• First test WLAN using another computer, to rule out

computer related malfunction.

• Check that WLAN has been enabled on the PC

• > Check that all relevant settings are set to enable WLAN

• > Check that WLAN is not disabled by the driver.

• Check that the WLAN dongle is set up with the correct

TCPIP settings. Default is using DHCP (Obtain an IP
address automatically).

• WLAN adapter drivers and utilities may come equipped

with features that allow intelligent control of the radio.
These features may automatically shut down the adapter.
This occurs typically after the computer has been in sleep
or hibernate mode. It may be necessary to disable such
features. Consider removing OEM utilities all together. Use
Windows built-in features instead.

WLAN CONNECTION

Problem

• WLAN Connection is lost.

Possible Solutions

• Bad Reception or connectivity when running a network of

manikins. Ensure that the WPA security keys are not the
same on all manikins

(channel and passphrase)

.

If WLAN is functioning on the PCs and you are unable to connect
to the manikin via wireless connection, this indicates a problem
with the WLAN dongle in the manikin.
• Check network settings using Manikin Update utility See

section Wireless and LAN configuration

.

• As a last resort you may restore the manikin to factory

defaults. Contact your local Laerdal customer support.

If the manikin is identifiable on the wireless settings of the PCs,
but cannot be connected to, the problem could be related to the
software security keys.

• SimMan Essential uses WPA authentication as default. The

default password is SimMan3G.

• Should a problem arise with the WLAN profiles, please

delete the existing WLAN profiles for the manikin and try
reconnecting to the manikin.

Check the wireless connection between the manikin and the PCs:
• Make sure that the manikin and the PC are ON.

• From the PC, Run command prompt: shortcut Windows-

button + R, type cmd and press enter.

• Typing: ping –t 192.168.170.1 (with wireless connection),

192.168.169.2 (with cable connection). The same numbers
apply for all manikins.

• Error messages are generated if connection is missing. Retry

or contact Laerdal support.

CHANGING MANIKIN WIRELESS NETWORK DISPLAY
NAME
If operating more than one manikin, ensure that each manikin has
it’s own unique SSID name. For more information see, section
Wireless and LAN configuration.

VOICE CONFERENCE APPLICATION

Problem

• Sound from manikin to Instructor not working.

Possible Solutions

• Check that the microphone has not slipped out of place.

Open the head skin zipper, located at the back of the head,
enough to reveal the ears.

• Ensure that the microphone is positioned in the cup located

towards the top of the ear with the black surface facing
outwards.