Troubleshooting – Laerdal SimMan Essential User Manual

Page 69

Advertising
background image

69

Manikin Setup

System Setup

Sim

ulation Setup

Clinical F

eatur

es

Maintenance

Tr

oubleshooting

Spar

e Par

ts

Troubleshooting

VOICE CONFERENCE APPLICATION

Problem

• Instructor microphone not picking up sound.

Possible Solutions

• Unplug and re-plug the headset to your computer

• Ensure correct sound device is selected. In Voice Conference

Application main menu, select <Options>, <Select Device>.

• Check volume settings in windows. Ensure microphone is

not muted.

THE DEBRIEF VIEWER

Problem

• Missing video capture - If the Web-camera does not record

video for debriefing.

Possible Solutions

• Check that the web-camera’s USB-cable is plugged into the

Patient Monitor PC.

• Check the web-camera settings via the Profile Editor.

Ensure settings match the web-camera setup.

• Ensure you use the correct profile file.

• Ensure that there is only one USB web-camera connected

to the PC.

• The video recordings will be stored on the PC connected

to the web camera. Ensure the computer connected to the
web-cam is available when transferring to debrief.

Problem

• Patient Monitor video capture is not included in the

debriefing.

Possible Solutions

• Check that the name of the Patient Monitor PC being used

corresponds with the setting in the profile in use.

MANIKIN

Problem

• Unpredictable behavior

Possible Solutions

• Manikin malfunctions may be caused by loose cables, tubes

or connectors. Open the torso and check if any items
appear to have become disconnected or are leaking. See
section: Manikin Setup - Opening the Torso.

• In case of fluid leakage, power off the manikin and contact

Laerdal Technical Service.

Problem

• Identifying a single manikin if multiple manikins in network

• When checking the manikin’s pulse, the <Select Manikin>

dialog of the Instructor Application will indicate which
manikin is pulse palpated.

AIRWAY CONTAMINATION

Problem

• Manikin airways have become contaminated from mouth-

to-mouth rescue breathing

Solution

• Clean the outside of the manikin with manikin wipes. Clean

the inside of the oral cavity with manikin wipes. Change
the lung bladders; see section: Maintenance - Replacing Lung
Bladders

. The manikin airways are not designed for mouth to

mouth rescue breathing or to be disinfected.

CHEST MOVEMENT

Problem

• No chest rise on manikin.

Solution

• Check that manikin is switched ON.

• Check that manikin is not in sleep mode due to inactivity.

Reactivate the manikin.

• Ensure that the condensation valve is closed. See section:

Opening the Condensation Valve.

• Check that awRR is not set to zero on Instructor

Application (Patient Case or scenario)

• Check that airway complications like maximum air resistance

or laryngospasm are not set.

• Check if the internal compressor is switched off. See

section: Turning the Internal Compressor Off/On.

• Check that any external compressed air source is switched

off and that the air tube is disconnected from the manikin.

• The internal compressor be overheated. Wait approximately

20 minutes for it to cool down.

• Chest-rise is set to bilateral (for example if ET-tube is

inserted to far into the bronchia).

• Chest-rise bladder is leaking or tubing to chest-rise bladder

is twisted kinked or disconnected. Replace chest-rise
bladder if it is leaking, see Maintenance section: Replacing
Chest Rise Bladders.

• Check air tubing for leakage; check that all connections are

intact. Replacing tubing section if leaking.

• Shallow chest movements and the internal compressor

runs continuously. The internal compressor may be worn -
contact Laerdal Technical Service.

MANIKIN LIMBS

Problems

• Lack of motion in the legs

Possible Solution

• Loosen and re-adjust the hip joint nuts on the inside of the

pelvis. See section: Manikin Setup - Attaching the legs.

Advertising