Troubleshooting - iv arm, Simulation setup – Laerdal SimMan Essential User Manual

Page 43

Advertising
background image

43

Manikin Setup

System Setup

Sim

ulation Setup

Clinical F

eatur

es

Maintenance

Tr

oubleshooting

Spar

e Par

ts

Simulation Setup

Troubleshooting - IV Arm

Problem

• Veins do not energize.

Solution:

• Ensure manikin is powered on and IV Arm cable and air tube are connected to the manikin.

• Ensure the blood reservoir is filled, see Filling and priming the system.

• Ensure the tourniquet area is activated (depressed).

• Look for a leak between the IV Arm pad and the IV Arm chassis.

• If the tourniquet area has been activated several times within a short period of time, internal pressure in the IV Arm may drop.

Wait 30 seconds to let pressure rebuild and try again.

Problem

• When injecting more than 40ml fluid and prime switch is depressed, no fluid is visible in the drain tube.

Solution

• Leak inside IV Arm. Contact Laerdal Technical Service.

Problem

• After releasing the prime switch and filling the internal reservoir, more than 16ml of fluid can be injected into the arm.

Solution

• Leak inside the IV Arm. Contact Laerdal Technical service

Problem

• There is a leak between IV Arm Pad and IV Arm chassis.

Solution

• Ensure the IV Arm Pad is fully inserted and secure in the IV Arm chassis.

• If this does not help, too high pressure has been applied to the IV Arm Pad during the filling process. Replace the IV Arm Pad.

Problem

• No flashback when inserting Intravenous Catheter into IV Arm Pad

Solution

• Ensure that the collection bag is at the same elevation as the IV Arm.