Troubleshooting

LUNGS

Problem

• Lungs not functioning properly

Possible Solution

• Check that the Airway resistance is not set to maximum in

the Instructor Application.

• Open the torso and chest plate. Check that the lungs are

free to expand and are not restricted by any cables.

• Check that the lung bladders are properly connected, and

that the tubes are not twisted.

• Check that the lung bladders are in a horizontal position

and inserted correctly. Ensure that the lung compliance
O-rings lie between the folds of the lung bladders.

• Check for flaws or ruptures in the lung bladders.

• Check that the two lung compliance O-rings are fitted

correctly. Replace O-rings if they appear to be damaged.

• Check that there are no obstructions inside the manikin

airways which may block air flow.

• If there is no change when adjusting lung compliance,

contact Laerdal Technical Service.

• If there is no change in lung resistance, contact Laerdal

Technical Service.

MECHANICAL NOISE DURING AUSCULTATION
In the Instructor Application, click <auscultation focus>.

BATTERIES

Problem

• Battery life is less than 150 minutes with healthy patient and

fully charged batteries.

Possible Solutions

• Batteries may be old, (recommended lifetime is 200

discharge cycles). Insert new batteries.

• Compressor may not function properly; consult your local

Laerdal Technical Services Centre.

PULSES

Problem

• Cannot feel pedal pulses

Possible Solutions

• Skin may be too tight over pulse units – re-adjust skin and

reboot.

CLINICAL FEATURES

-

PNEUMOTHORAX BLADDERS

Problem

• Pneumothorax bladders are not functioning properly.

Solution

• Check tubing connection at base of the bladder, to ensure

that tubing has not been disconnected.

MANIKIN SHUTDOWN

Problem

• The manikin is unresponsive.

Solution

• Press and hold the <ON/OFF> button for 10 sec to force

shutdown the manikin.