Clinical Features

3 - LEFT LUNG RESISTANCE
The On setting will cause a full airway obstruction on the manikin.
• Off - Normal

• On – Extreme

Off corresponds to Resistance = 0 in scenario editor.
On corresponds to Resistance = 3 in scenario editor.

4 - ALLOW STOMACH DISTENTION
Check or uncheck the box to activate stomach distention or uncheck the box to allow trapped air to exit.

5 - EXHALE CO

2

Check or uncheck the box to enable or disable CO

2

exhalation. (Requires connection to an external CO

2

reservoir)

6 - TONGUE FALLBACK
Check the box to simulate tongue fallback. Uncheck the box to disable this function. Tongue fallback symptoms will occur unless head
is tilted back or jaw thrust maneuver is applied.

Breathing Features

Warning: Do not ventilate the manikin with oxygen enriched air or flammable gasses.

The SimMan Essential can simulate spontaneous breathing and can be connected to ventilators or used with assisted ventilations:

• Bilateral and unilateral chest rise and fall

• Normal and abnormal breath sounds

• 5 anterior auscultation sites and 6 posterior auscultation sites

• Unilateral, bilateral and lobar breath sounds

• Oxygen saturation and phlethysmogram

• CO

2

exhalation for use with third-party et CO

2

detectors (Requires connection to an external CO

2

reservoir)

PATIENT MONITOR FEATURES - BREATHING (OPPTIONAL)
• SpO

2

• Airway respiration rate (awRR)

• End-tidal CO

2

(etCO

2

)

• End-tidal O

2

(etO

2

)

• inO

2

• PH

LUNG SPECIFICATIONS

• Max tidal volume: 1.2 liters

• Max tidal volume registered in the Instructor Application is 900ml. All volumes higher than 900ml will register as 900ml

• Max airway pressure: 80 cm H

2

O

• Simulated stomach inflation starts from approximately 40cm H

2

O airway pressure. Lungs are not intended for use with

PEEP-valves.