Simulation Setup

Air/CO

2

The Air / CO

2

is located on the left side of the torso. To access the panel, lift up the manikin skin flap and remove the protective

covering. Connect external Air / CO

2

inlet.

SIMMAN ESSENTIAL PANEL

CONNECTING EXTERNAL AIR AND CO

2

SUPPLY

An internal compressor is located in the right leg of the manikin. It is recommended to use an external source of compressed air
whenever the manikin is stationary over extended periods of use.

If the manikin is required to exhale CO

2

with each ventilation - connect external CO

2

.

Connect CO

2

only if the manikin is required to exhale CO

2

with each ventilation. Exhaled CO

2

can be detected with a real capnographic

device.

1 Connect a suitable CO

2

source to a Laerdal external compressor or regulator panel.

2 Connect a Laerdal double-lumen Air/CO

2

tube from the external compressor or regulator panel to the Air/CO

2

inlet on the panel.

For more information on external compressors and regulator panels compatible with SimMan Essential, contact your local Laerdal representative.