9. Read Clock

P:

#np0000RCL^J

R:

#np16/06/2011 10:05:23

10. Write Clock

P:

#np0000WCL16/06/2011 10:05:23^J

R:

#npACK ó #npERR

11. Write Nominal Voltage and Nominal Frequency

P:

#np0000WVH23000 500^J

R:

#npACK ó #npERR

12. Read Base battery disconnection time

P:

#np0000RTB^J

R:

#np0600

13. Delete Positive Energies

P:

#np0000CLEP^J

R:

#npACK ó #npERR

14. Delete Negative Energies

P:

#np0000CLEN^J

R:

#npACK ó #npERR

15. Reset MaXimum Values

P:

#np0000RMX^J

R:

#npACK

16. Reset MiNimum Values

P:

#np0000RMN^J

R:

#npACK

17. Write Peripheral Number

P:

#np0000WNP05^J

R:

#npACK ó #npERR

11.5.- FTP

QNA500 has an internal FTP server that is used to store the files with the data measured by the power

quality analyzer. These files can be downloaded with a standard FTP connection from a PC. The FTP

servers have been specially designed for the transmission or downloading of files. This achieves higher

transmission or data download speeds than with the standards used in serial ports. Likewise, it

facilitates the development of third party applications to download files from the analyzer and

subsequently manage this information.