Calibrate:transient:offset, Calibrate:value:current, Calibrate:value:voltage – AMETEK PLA-PLW Programming Manual User Manual

Page 18

Advertising
background image

16

calibration constants are lost unless they have been stored with CALibrate:SAVE.

Command Syntax

CALibrate:STATe <bool>

Parameters

0 | 1 | OFF | ON

*RST Value

OFF

Query Syntax

CALibrate:STATe?

Returned Parameters

<NR1>

CALibrate:TRANsient:OFFSet

This command can only be used in calibration mode. It is used to set the DAC count of
immediate level when enter transient calibration mode.

Command Syntax

CALibrate:TRANsient:OFFSet <NR1>

Parameters

<NR1> DAC count of immediate level

Query Syntax

CALibrate:TRANsient:OFFSet?

Returned Parameters

<NR1>

CALibrate:VALue:CURRent

This command is only used in calibration mode. It enters a calibration current that you obtain by
reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value
being entered. These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.

Command Syntax

CALibrate:VALue:CURRent <NRf>

Parameters

<NRf> current value

CALibrate:VALue:VOLTage

This command is only used in calibration mode. It enters a calibration voltage that you obtain by
reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value
being entered. These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.

Command Syntax

CALibrate:VALue:VOLTage <NRf>

Parameters

<NRf> voltage value