SYSTem:EXTernal:CONTrol
This command sets or queries the syntax mode state
Command Syntax
SYSTem:EXTernal:CONTrol <NR1>
Parameters
<NR1> Toggles External Control State On/Off
0 = External Control Off
1 = External Control On
Query Syntax
SYSTem:EXTernal:CONTrol?
Parameters
None
Returned Parameters
<NR1> integer value representing external control state
SYSTem:GPIB:ADDRess
This command sets or queries the value of the GPIB address.
Note: The new GPIB address will only take effect after power cycle.
Command Syntax
SYSTem:GPIB:ADDRess <NR1>
Parameters
<NR1> GPIB address, 1 to 30
Query Syntax
SYSTem:GPIB:ADDRess?
Parameters
None
Returned Parameters
<NR1> integer value representing GPIB address
SYSTem:GPIB:EOS
This command sets or queries the value of the end-of-string (EOS) code for GPIB interface.
Command Syntax
SYSTem:GPIB:EOS <NR1>
Parameters
<NR1> end-of-string(EOS) code
0 = NULL(no EOS)
1 = CR
2 = LF
3 = CR/LF
Query Syntax
SYSTem:GPIB:EOS?
Parameters
None
Returned Parameters
<NR1> integer value representing EOS code
SYSTem:INPut:BOOT
The command sets and queries the state of input when syst boot/init.
Command Syntax
SYSTem:INPut:BOOT <NR1>
Parameters
<NR1>
0 = input is OFF
1 = input is ON
Query Syntax
SYSTem:INPue:BOOT?
Parameters
None
Returned Parameters
<NR1> integer value representing system boot state.
SYSTem:MODel
The command queries the model specification.
Query Syntax
SYSTem:MODel?
Parameters
None
Returned Parameters
<AARD> aaaW-bbbV-cccA-dddKOHM