System:net:subnet, System:range, System:serial – AMETEK PLA-PLW Programming Manual User Manual

Page 53: System:short:status, System:version

Advertising
background image

49

Returned Parameters

<AARD>

Four

fields: <serial> <dhcp> <ip address> <idn string>

<serial> serial number, same as SYST:SERial?

<dhcp>

assigned flag of DHCP, same as SYST:NET:DHCP?

<ip

address> IP address, same as SYST:NET:ADDR?

<idn

string> identification string, same as *IDN?

SYSTem:NET:SUBNet

The command sets or queries the device subnet mask.

Command Syntax

SYSTem:NET:SUBNet <AARD>

Parameters

<AARD> decimal dot notation.

Query Syntax

SYSTem:NET:SUBNet?

Returned Parameters

<AARD> string representing subnet mask.

SYSTem:RANGe

This command queries the range number for all four operational modes.

Query Syntax

SYSTem:RANGe?

Returned Parameters

<AARD> nn1/nn2/nn3/nn4

nn1 = range number of constant voltage

nn2 = range number of constant current

nn3 = range number of constant resistance

nn4 = range number of constant power

SYSTem:SERial?

This command queries the serial number of the main control board.

Query Syntax

SYSTem:SERial?

Parameters

None

Returned Parameters

<CRD> SN:000B87XXXXXX

SN:000B87 = Fixed string

XXXXXX

=

serial number of main control board

SYSTem:SHORT:STATus?

This command queries the serial number of the main control board.

Query Syntax

SYSTem:SHORt:STATus?

Parameters

<NR1> Toggle on/off

0 = OFF

1 = ON

Returned Parameters

<NR1> integer value representing short status state

SYSTem:VERSion?

This query returns the firmware version. The value is of the form FVX.XX, where X.XX is the revision
number.

Query Syntax

SYSTem:ERSion?

Parameters

None

Returned Parameters

<CRD> string representing version number

Advertising