Source:]current:slew:negative, Source:]current:slew[:positive, Source:]current:tlevel, itr – AMETEK PLA-PLW Programming Manual User Manual

Page 22

Advertising
background image

20

slew time value may be entered, the electronic load selects a slew time that is closest to the
programmed value. MINimum sets the slew to the fastest possible time. MAXimum sets the slew to
the slowest time.

Note: Slew time greater than the maximum value is set to MAXimum. Slew time less than the
minimum value are set to MINimum.

Command Syntax

CURRent:SLEW:BOTH <NRf+>

Parameters

Slew time. Refer to model specifications | MIN | MAX

Unit

ms

Query Syntax

None

[SOURce:]CURRent:SLEW:NEGative

This command sets the slew time of the current for negative going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.

Note: Slew time greater than the maximum value is set to MAXimum. Slew time less than the
minimum value are set to MINimum.

Command Syntax

CURRent:SLEW:NEGative <NRf+>

Parameters

Slew time. Refer to model specifications | MIN | MAX

Unit

ms

Query Syntax

CURRent:SLEW:NEGative?
CURRent:SLEW:NEGative? MIN

CURRent:SLEW:NEGative? MAX

Returned Parameters

<NRf+>, slew time.

[SOURce:]CURRent:SLEW[:POSitive]

This command sets the slew time of the current for positive going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.

Note: Slew time greater than the maximum value is set to MAXimum. Slew time less than the
minimum value are set to MINimum.

Command Syntax

CURRent:SLEW <NRf+>
CURRent:SLEW:POSitive <NRf+>

Parameters

<NRf+>: slew time. Refer to model specifications | MIN | MAX

Unit

ms

Query Syntax

CURRent:SLEW?
CURRent:SLEW? MIN

CURRent:SLEW? MAX

CURRent:SLEW:POSitive?
CURRent:SLEW:POSitive? MIN
CURRent:SLEW:POSitive? MAX

Returned Parameters

<NRf+>, slew time

[SOURce:]CURRent:TLEVel, ITR