Beurer GL50 evo mg/dL

Note

The blood glucose monitor is made of precision components. The accuracy of the measurements

and service life of the device depend on its careful handling:

• Protect the device from impacts and do not drop it.

• Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or

water, marked temperature fluctuations, direct sunlight and extreme cold.

• Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile

telephones.

8 W

hat

if

there

are

problems

?

Display messages on batteries and blood glucose measurement

No.

Cause

Solution

Empty

battery

Batteries discharged.

Charge the blood glucose monitor via the USB

interface.

High

temperature

Very high glucose level

(> 630mg/dL)

Temperature of the measuring environment,

blood glucose monitor or test strip above the

permitted range.

Repeat the test using a new test strip as

soon as the measuring environment, blood

glucose monitor and test strips have reached

room temperature (+20°C to +26°C).

Low

temperature

Very low glucose level

(< 20mg/dL)

Low temperature Temperature of the measur-

ing environment, blood glucose monitor or test

strip below the permitted range.

Repeat the test using a new test strip as soon

as the measuring environment, blood gluco-

se monitor and test strips have reached room

temperature (+20°C to +26°C).

Strip
error

Used or contaminated test strip

inserted.

Insert unused test strip that has not expired.

Repeat the blood glucose measurement.

Error 1

System error

Remove batteries, reinsert batteries.

Contact our customer service if the problem

persists.

Error 2

Insufficient amount of blood on

the test strip.

Repeat the measurement using a new test

strip.