Beurer GL 50 evo (mgdl) User Manual

Page 33

Advertising
background image

Beurer GL50 evo mg/dL

33

Problem: device does not switch on

Cause

Solution

Battery empty.

Charge the blood glucose monitor via the USB

port on a computer.

Test strip inserted incorrectly or not completely. Firmly insert the test strip into the slot on the

device with the contacts first. Make sure that

the front of the test strip is facing you (see “Test

strip” on page 13).

Device faulty.

Contact Customer Services.

The sliding switch is in the “off” position

To activate the device set the sliding switch to

setting “on”

Problem: the test does not start after inserting the test strip into the device and applying

blood.

Cause

Solution

Insufficient blood or test strip not filled cor-

rectly.

Repeat test using a new test strip and a larger

drop of blood.

Faulty test strip.

Repeat the test using a new test strip.

Blood was applied while the device was

switched off.

Repeat the test using a new test strip and only

apply blood when finger is shown on display.

The basic settings of the device have been

changed and these changes were not com-

pleted (see “4.2 Making and changing basic

settings” on page 15).

Remove the test strip and press the “scroll

wheel” button until the device is switched off.

Device faulty.

Contact Customer Services.

Error 5

System error

Remove batteries, reinsert batteries.

Contact our customer service if the problem

persists.

Unknown error messages.

Remove batteries, reinsert batteries. Contact

Customer Services if the problem persists.