11 u – Beurer GL 50 evo (mgdl) User Manual

Page 37

Advertising
background image

Beurer GL50 evo mg/dL

37

Results of the system accuracy for combined glucose concentrations between 34.4 mg/dL

(1.9 mmol/L) and 442.8 mg/dL (24.6 mmol/L).

Within ±15 mg/dL or ±15%

(Within ±0.83 mmol/L or ±15%)

597/618 (96,6 %)

In comparison to the YSI, the GL50 evo met the EN ISO 15197:2013 standard, whereby 95% of the

blood glucose values measured have to fall within the following zones: either ±0.83 mmol/L (±15 mg/

dL) of the measured average value when using the reference measuring procedure for blood glucose

concentrations <100 mg/dL (<5.55 mmol/L) or ±15% for blood glucose concentrations of ≥100 mg/

dL (≥5.55 mmol/L). 99% of the individual measured blood glucose values must fall within zones A

and B of the Consensus Error Grid (CEG) for diabetes type 1.

Performance evaluation by the user
A study to assess the glucose values of blood samples of capillary blood from the fingertips, which

were obtained from 103 individuals that had no special training, produced the following results:

96.7% within ±15 mg/dL (±0.83 mmol/L) and 95.9% within ±15% of the values obtained in the me-

dical laboratory with glucose concentrations of at least 100 mg/dL (5.55 mmol/L).
You will find further details and information regarding blood glucose results and various technologies

in generally relevant specialist medical literature.

11 u

sage

limits

for

specialist

personnel

from

the

healthcare

sector

:

1. If the patient shows the following symptoms, it may be the case that no correct values can be

displayed:

• Acute dehydration

• Acute hypotension (low blood pressure)

• Shock

• Hyperosmolar hyperglycaemic condition (with or without ketosis)

2. Lipaemic samples: cholesterol levels up to 500 mg/dL and triglycerid levels up to 1000 mg/dL do

not influence the results. Severely lipaemic blood samples were not tested with the Beurer GL50

evo blood glucose monitor; therefore, using the device with these samples is not recommended.

3. In the case of severely ill patients, blood glucose monitors for home use should not be used.

4. The influence of interfering substances on the measurements depends on the concentration in

the blood. The maximum concentrations of certain substances listed below do not significantly in-

fluence the measurements.

Advertising