S T A N D A R D S A N D P R I N C I P L E S
8 - 1 5
Test Principle: Protective Earth Resistance
Notes:
S1 = Not applicable
S2 = Not applicable
S3 = Open
S5 = Open
baw137f.eps
Function:
Measures the impedance between the 601PRO’s outlet PE terminal and exposed conductive parts of
the DUT that are connected to the DUT’s protective earth. Test current is applied for approximately 5
seconds.
Applicable to:
IEC 601-1, Class I:
Current source = 25 A ± 10%
VDE 751-1, Class I:
Current source = 5-25 A
HEI 95, Class I:
Current source = 1 A
VDE 701, Class I:
Current source = 1 A
IEC 1010, Class I:
Current source = 1 A
AAMI:
Current source = 1 A
AS/NZS 3551:
Current source = 1 A