Computer control commands (cont.) – Fluke Biomedical 601PRO XL User Manual
Page 153

C O M P U T E R C O N T R O L
1 0 - 5
Computer Control Commands (Cont.)
Voltage Measurement Commands
COMMAND DESCRIPTION
NOTES
[VL1] Voltage
L1-EARTH
[VL2] Voltage
L2-EARTH
[V12] Voltage
L1-L2
DUAL Mode Measurements
COMMAND DESCRIPTION
NOTES
[DLV]
Dual Lead Voltage
Current Measurement Commands
COMMAND DESCRIPTION
NOTES
[CUR] Current
Consumption
Insulation Resistance Measurement Commands
COMMAND DESCRIPTION
NOTES
[IMN] Mains-Case
[IAL]
ALL - Case
[IRA]
RA - Case
[IRL]
RL - Case
[ILA]
LA - Case
[ILL]
LL - Case
[IV1] V1-V6-Case
Protective Earth Resistance Measurement Commands
COMMAND DESCRIPTION
NOTES
[PEC]
Protective Earth Resistance
[CAL]
Calibration of PEC at selected current
[S01]
Select 1 Amp
[S10]
Select 10 Amp
[S25]
Select 25 Amp