Ecg output – Fluke Biomedical 601PRO XL User Manual
Page 77

M A N U A L M O D E
3 - 3 7
ECG Output
The 601PRO generates a series of ECG, performance, and arrhythmia waveforms to
verify the accuracy of ECG machines/monitors. ECG Output can ONLY be accessed
from the
Manual
menu.
The following outlet condition applies:
⇒ Normal Polarity, Earth Open, Outlet ON
To generate waveforms:
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Attach the device’s ECG leads to the matching 601PRO applied part terminals.
• From the
MAIN MENU
, press
TESTS/AUTOMODES
, then
MANUAL
, then
ECG OUTPUT
.
• At the
ECG/Performance Waveforms
menu, press the
SOFT KEY
below the
type of waveform/frequency to output. The outlet is powered on.
baw066f.eps
• For example, press
SOFT KEYS 1 - 3
to choose from the waveforms.
baw067f.eps
• Press
SOFT KEY 4 (
*
MORE)
for additional waveforms.
Sample waveforms are shown on the following pages.