Fluke Biomedical 601PRO XL User Manual
Page 75

M A N U A L M O D E
3 - 3 5
To perform a VDE Equivalent Patient Leakage test:
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Attach Patient Leads to 601PRO ECG jacks.
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Press the
-
shortcut key
. The
Equivalent Patient Leakage
test is displayed:
baw064f.eps
• Press
START TEST (SOFT KEY 1)
to begin the test.
WARNING:
High voltage is present at the Applied Parts terminals while
measurements are being taken.
• Press the
print data
key to generate a printout of the latest measurement.
To perform a calibration from the
Equivalent Patient Leakage
test,
press
CALIBRATION (SOFT KEY 2)
.
• Follow the same process described in
Equivalent Device Leakage
.