Description of the MCA527 Firmware Commands

Command name

CMD_SET_THRESHOLD

Compatibility

Identical to the MCA166 command.

Execution right

Necessary

Description

This command sets the threshold value (thr: 0 … 60 percent).

Format

integer

integer

integer

long

integer

Parameter

preamble

command

thr

0

end flag

Byte string (HEX)

A5

5A

47

00

thr

00

00

00

00

00

B9

9B

Remarks

See next command.

Command name

CMD_SET_THRESHOLD_TENTHS

Compatibility

New MCA527 command.

Execution right

Necessary

Description

This command sets the threshold value (thr: 0 … 600 * 0.1 percent).

Format

integer

integer

integer

long

integer

Parameter

preamble

command

thr

0

end flag

Byte string (HEX)

A5

5A

0D

01

thr

l

thr

h

00

00

00

00

B9

9B

Remarks

This command has been added for setting the threshold with higher resolution.

Command name

CMD_SET_SHAPING_TIME

Compatibility

Compatible to the MCA166 command.

Execution right

Necessary

Description

The command sets the shaping time (dtc).
dtc = 1 shaping time low
dtc = 3 shaping time high
The command is ignored and responds with an error value if a measurement is still running.

Format

integer

integer

integer

long

integer

Parameter

preamble

command

dtc

0

end flag

Byte string (HEX)

A5

5A

52

00

dtc

00

00

00

00

00

B9

9B

Remarks

The MCA527 provides more than two shaping times, but in order to keep compatibility with older
software, this command has been retained. The command allows furthermore to chose the shaping
time from a pair of shaping times. This pair can be set by CMD_SET_SHAPING_TIME_PAIR.

Command name

CMD_SET_SHAPING_TIME_PAIR

Compatibility

New MCA527 command.

Execution right

Necessary

Description

The command sets the shaping time pair.(lst = 1 … 254 * 0.1 µs, hst = 2 … 255 * 0.1 µs, lst<hst).
The highest allowed shaping time is returned by CMD_QUERY_STATE527, byte offset 121.
The command is ignored and responds with an error value if a measurement is still running.

Format

integer

integer

integer

integer

integer

integer

Parameter

preamble

command

lst

hst

0

end flag

Byte string (HEX)

A5

5A

0C

01

lst

00

hst

00

00

00

B9

9B

Remarks

See also previous command.

11