Description of the MCA527 Firmware Commands

Command name

CMD_QUERY_ENRICHMENT

Compatibility

Identical to the MCA166 command.

Execution right

Not necessary

Description

The command reads current dead time, real time and the integral of two ROIs. This is very useful
during a running measurement because in this way there is no need for transferring parts of the
spectrum data for calculating the ROI integrals itself. The name of the command refers to its original
usage. The command is, of course, usable for all applications that need this information.

r1b:

ROI

1

begin

r2b:

ROI

2

begin

r1l:

ROI

1

length (8 ... 250 channels)

r2l:

ROI

2

length (8 ... 250 channels)

Format

integer

integer

char

char

integer

integer

integer

Parameter

Preamble

command

r1l

r2l

r1b

r2b

end flag

Byte string (HEX)

A5

5A

5D

00

r1l

r2l

r1b

l

r1b

h

r2b

l

r2b

h

B9

9B

Result data array

Byte offset 0

Dead time

unsigned long

Byte offset 4

Real time

unsigned long

Byte offset 8

Integral of the 1

st

ROI

unsigned long

Byte offset 12

Integral of the 2

nd

ROI

unsigned long

Byte offset 16

unused

80 bytes

Byte offset 106

Command flag and parameters

8 bytes

Byte offset 114

unused

12 bytes

Byte offset 126

Checksum

unsigned short

Byte offset 128

unused

4 bytes

47