Cmd_query_adjustment_table – GBS Elektronik MCA-527 Firmware Commands User Manual

Page 54

Advertising
background image

MCA527

Name

CMD_QUERY_ADJUSTMENT_TABLE

Compatibility

New MCA527 command.

Execution right

Not necessary

Description

The command returns the ADC values that has been measured with different settings at open signal
input. These data can be used to interpolate

the ADC value which corresponds to the ground

line. A value is valid if it is unequal to -1.

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

0

0

end flag

Byte string (HEX)

A5

5A

26

01

00

00

00

00

00

00

B9

9B

Result data array

Byte offset 0

measured with coarse gain = 2, offset DAC = 1024

short

Byte offset 2

measured with coarse gain = 2, offset DAC = 15360

short

Byte offset 4

measured with coarse gain = 5, offset DAC = 1024

short

Byte offset 6

measured with coarse gain = 5, offset DAC = 15360

short

Byte offset 8

measured with coarse gain = 10, offset DAC = 1024

short

Byte offset 10

measured with coarse gain = 10, offset DAC = 15360

short

Byte offset 12

measured with coarse gain = 20, offset DAC = 1024

short

Byte offset 14

measured with coarse gain = 20, offset DAC = 15360

short

Byte offset 16

measured with coarse gain = 50, offset DAC = 1024

short

Byte offset 18

measured with coarse gain = 50, offset DAC = 15360

short

Byte offset 20

measured with coarse gain = 100, offset DAC = 1024

short

Byte offset 22

measured with coarse gain = 100, offset DAC = 15360

short

Byte offset 24

unused

82 bytes

Byte offset 106

Command flag and parameters

8 bytes

Byte offset 114

unused

12 bytes

Byte offset 126

Checksum

unsigned short

Byte offset 128

MCA state

unsigned short See CMD_QUERY_POWER

Byte offset 130

unused

2 bytes

54