Description of the MCA527 Firmware Commands

MCA Data Setup Commands:

Command name

CMD_SET_TDF

Compatibility

Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software

Execution right

Necessary

Description

The command does nothing. It has only been retained for compatibility with older software

Format

integer

integer

integer

long

integer

Parameter

preamble

command

tdf

0

end flag

Byte string (HEX)

A5

5A

61

00

tdf

l

tdf

h

0

0

0

0

B9

9B

Remarks

The command is meaningless for the current firmware version. Possibly, future firmware will use the
parameter again.

Command name

CMD_SET_UF6_ROIS

Compatibility

Compatible to the MCA166 command.

Execution right

Necessary

Description

The command sets the begin and the end of a ROI used by the other UF6 commands. The name of
the command refers to its original usage.
r:

ROI number (1, 2 or 3)

b:

ROI begin

e:

ROI end

Up to firmware version 14.01: b < e < max. channels number, (e − b) < 512.
Since firmware version 14.02: b < e < max. channels number

If b = 0 and e = 0, the corresponding ROI is reset to “not used”.

Format

integer

integer

integer

integer

integer

integer

Parameter

Preamble

command

r

b

e

end flag

Byte string (HEX)

A5

5A

64

00

r

00

b

l

b

h

e

l

e

h

B9

9B

Command name

CMD_SET_USER_DATA

Compatibility

Compatible to the MCA166 command.

Execution right

Necessary

Description

The command stores any 32 bit value (val) in the MCA memory (256 entries).
e: 0 ... 255

entry number

Format

integer

integer

integer

32 bit

integer

Parameter

preamble

command

e

val

end flag

Byte string (HEX)

A5

5A

57

00

e

00

...

...

...

B9

9B

Remarks

See also document “Use of MCA User Data Memory by Specific Applications”.