MCA527

Command name

CMD_SET_PUR

Compatibility

Identical to the MCA166 command.

Execution right

Necessary

Description

The command sets the pile up rejection.
pur = 0 turns PUR off
pur

0 turns PUR on

The command is ignored and responds with an error value if a measurement is still running.

Format

integer

integer

integer

long

integer

Parameter

preamble

command

pur

0

end flag

Byte string (HEX)

A5

5A

53

00

pur

00

00

00

00

00

B9

9B

Command name

CMD_SET_FAST

Compatibility

Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software

Execution right

Necessary

Description

This command sets the fast discriminator threshold (0 ... 2499, default 400).

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

fast

0

end flag

Byte string (HEX)

A5

5A

50

00

fast

l

fast

h

00

00

00

00

B9

9B

Command name

CMD_SET_SLOW

Compatibility

Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software

Execution right

Necessary

Description

This command sets the slow discriminator threshold (0 ... 2499, default 400).

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

slow

0

end flag

Byte string (HEX)

A5

5A

51

00

slow

l

slow

h

00

00

00

00

B9

9B

14