Cmd_set_bias, Cmd_set_pin5_current_source – GBS Elektronik MCA-527 Firmware Commands User Manual

Page 18

Advertising
background image

MCA527

Command name

CMD_SET_BIAS

Compatibility

Compatible to the MCA166 command.

Execution right

Necessary

Description

The command sets the detector high voltage and controls the HV inhibit input.
hv:

0 ... 3600 V

i = 0

Inhibit off

i = 1

”Canberra HPGe mode”,

HV shut down if inhibit input < 0.5V

20

i = 2

“DSG HPGe mode”,

HV shut down if inhibit input < 0.5V

20

i = -1

”Ortec HPGe mode”,

HV shut down if inhibit input ≥ 5V

Format

integer

integer

integer

integer

integer

integer

Parameter

preamble

command

hv

i

0

end flag

Byte string (HEX)

A5

5A

4F

00

hv

l

hv

h

i

l

i

h

00

00

B9

9B

Remarks

The Lite version provides no HV inhibit input. Setting the HV inhibit input on is ignored.
The OEM version provides neither HV nor HV inhibit input. All settings are ignored.
If the connected detector supplies detector information, the MCA527 compares the polarity of the
detector with that of the power module. If the polarities are different, all settings are ignored.

Command name

CMD_SET_PIN5_CURRENT_SOURCE

Compatibility

New MCA527 command

Execution right

Necessary

Description

The command switches the current source (cs) on SUB-D9 pin5 on (= 1) or off (= 0).

Format

integer

integer

integer

long

integer

Parameter

preamble

command

cs

0

end flag

Byte string (HEX)

A5

5A

19

01

cs

00

00

00

00

00

B9

9B

Remarks

The SUB-D9 pin5 is supported by the full version and partially by OEM version.

20 Mode 1 and 2 are identical. The difference is made by reasons of legacy.