Cmd_query_uf6_rois – GBS Elektronik MCA-527 Firmware Commands User Manual

Page 48

Advertising
background image

MCA527

Name

CMD_QUERY_UF6_ROIS

Compatibility

Identical to the MCA166 command.

Execution right

Not necessary

Description

The command reads begin and end of up to three ROIs that can be set by the command
CMD_SET_UF6_ROIS. The name of the command refers to its original usage. See also the
command CMD_QUERY_UF6_INFO.

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

0

0

end flag

Byte string (HEX)

A5

5A

65

00

00

00

00

00

00

00

B9

9B

Result data array

Byte offset 0

1

st

ROI begin

unsigned long

Byte offset 4

1

st

ROI end

unsigned long

Byte offset 8

2

nd

ROI begin

unsigned long

Byte offset 12

2

nd

ROI end

unsigned long

Byte offset 16

3

rd

ROI begin

unsigned long

Byte offset 20

3

rd

ROI end

unsigned long

Byte offset 24

unused

82 bytes

Byte offset 106

Command flag and parameters

8 bytes

Byte offset 114

unused

12 bytes

Byte offset 126

Checksum

unsigned short

Byte offset 128

unused

4 bytes