MCA527

Command name

CMD_QUERY_STATE (2

nd

Continuation)

Result data array

Byte offset 100

Start time

unsigned long See CMD_START

Byte offset 104

TDF

unsigned short always 0 at MCA527

Byte offset 106

Command flag and parameters

8 bytes

Byte offset 114

Read out buffer state

unsigned short OCCUPIED

= 0x2000

OVERRUN

= 0x4000

FILLED

= 0x8000

Byte offset 116

Counts per second

35

unsigned long

Byte offset 120

Differential dead time [ ‰ ]

unsigned short

Byte offset 122

HV inhibit mode

short

Byte offset 124

HV inhibit state

unsigned short

Byte offset 126

Checksum

unsigned short

Byte offset 128

MCA state

unsigned short See CMD_QUERY_POWER

Byte offset 130

Start flag

36

unsigned short See CMD_START