Description of the MCA527 Firmware Commands

MCA Setup Commands:

Command name

CMD_SET_ADC_RES_DISCR

Compatibility

Compatible to the MCA166 command.

Execution right

Necessary

Description

The command sets the ADC resolution (res: 128, 256 … 16384 ), the LLD (Low level discriminator)
and the ULD (Upper level discriminator) to the values of its three parameter.
The maximum resolution is returned by CMD_QUERY_STATE527, byte offset 56.
The LLD must be smaller than the ULD. The maximum ULD is equal ADC resolution minus 1.
The command is ignored and responded with an error value if one or more parameter are invalid or a
measurement is still running.

Format

integer

integer

integer

integer

integer

integer

Parameter

preamble

command

res

LLD

ULD

end flag

Byte string (HEX)

A5

5A

46

00

res

l

res

h

LLD

l

LLD

h

ULD

l

ULD

h

B9

9B

Command name

CMD_SET_PRESETS

Compatibility

Compatible to the MCA166 command.

Execution right

Necessary

Description

The command sets the automatic stop condition (pre) to
PRESET_NONE

= 0

(val is meaningless)

PRESET_REAL

= 1

PRESET_LIVE

= 2

(val ≤ 65535)

13

PRESET_INT

= 3

PRESET_AREA

= 4

PRESET_REAL_MILLISECONDS

14

= 5

and value (val) of the preset.

Format

integer

integer

integer

long

integer

Parameter

preamble

command

pre

val

end flag

Byte string (HEX)

A5

5A

48

00

pre

l

pre

h

val

l

...

...

val

h

B9

9B

Remarks

Except for PRESET_LIVE

13

and PRESET_REAL_MILLISECONDS, the MCA527 stops the

measurement always at integer real time.
For general mode 3 and 4 (see CMD_SET_GENERAL_MODE), only PRESET_REAL is supported.
PRESET_LIVE, PRESET_INT and PRESET_AREA are without effect and mean the same like
PRESET_NONE.

13 Up to firmware version 13.03, the measurement was stopped always at integer seconds real time. Since firmware version 13.04,

if preset is equal to PRESET_LIVE, the measurement is stopped according to the preset live time and the resulting real time is
no more integer. The fractional digits are returned by CMD_QUERY_STATE527_EX, byte offset 80. Besides, in consequence of
the more exact automatic stop, the maximum value for PRESET_LIVE has been limited.

14 Since firmware version 14.03.