8. What if ...

This chapter will help you to recognize malfunctions and take
appropriate measures or, if possible and acceptable, to perform
some simple repairs.

8.1 Error messages

The furnace continuously checks all
functions during operation. If an
error is detected, the respective
error message is displayed.

The following error messages may be displayed. If there are any questions, please contact the Ivoclar Vivadent After Sales Service.

Error /

Hint

No.

Continua-

tion pos-

sible

Error

Error Message Text

2

T < B

Enter a logical value for T.

8

L > T

Enter a logical value for long-term cooling L.

9

V2x <= V1x

Enter a logical value for the vacuum-on temperature Vx1 or the vacuum-off temperature Vx2.

10

V2x > Tx + 1°C

Enter a logical value for V1x, V2x

11

Incorrect values for V1x, V2x

Geben sie plausible Werte für V1x, V2x ein.

13

*,**

Current temperature after Start

> Tx + 80°C

Excess temperature! Program aborted, furnace head opens to allow the furnace to cool down!

16

T2 < T1

Enter a lower value for T1 or a higher value for T2.

17

Power failure > 10 s during a

firing program in progress

A firing program in progress was interrupted for more than 10 s. The program cannot be continued!

18

T1 > V12

Enter a lower value for T1 or a higher value for V12.

19

vV set, but V2 is missing or

invalid

Pre-vacuum activated! V2 must be higher than B.

20

**

no

Error in the heating system

Check the heater fuse. If the fuse is i.o. contact your Service Center.

23

Heating muffle very old

The heating muffle is very old. It is recommended to replace it. After the error message has been

acknowledged, a firing program may still be started.

24

Heating muffle defective

The condition of the muffle is so poor that is has to be replaced immediately.

26

T is > B + 160 °C at the start of

a firing program

Firing chamber too hot to start a firing program.

27

**,***

no

Furnace head cannot be

initialized

The furnace head cannot be moved to the final position. It might be blocked by an external

mechanical source! If not contact your Service Center!

28

**

The furnace head does not

reach the target position

The furnace head does not open/close correctly. The furnace head was manually moved or is

obstructed. The furnace head must only be moved using the keys intended for this purpose!

32

**

no

The vacuum is not released

The vacuum cannot be released. The vacuum valve might be dirty or stuck. Contact your Service

Center.

33

Necessary vacuum (xxxmbar) is

not reached within 1 min.

The vacuum cannot be established. Check the seal of the firing chamber, vacuum hose, vacuum

pump, pump fuse.

103

Program start blocked

Starting a program is not possible due to a technical malfunction.

107

Incorrect time settings (date /

time)

The clock settings are incorrect. Please set a correct date and a correct time!

110

HV > H (H2)

Enter a lower value for HV or a higher value for H (H2).

120

„Share of the holding time with

vacuum“ is activated, but Vx2

does not correspond to Tx or

Tx+1

Activate the vacuum during the holding time Tx or deactivate HV.

500

Error pressing time

Max. pressing time exceeded.

504

Error press position

Max. position exceeded.

505

Error press force

Max. press force exceeded.

513

Error press drive initalized

Press drive is not initialized. Please switch the furnace off and on again.

514

Error press drive

Technical error in press drive.