2 technical malfunctions, 3 repair, 4 load factory settings – Ivoclar Vivadent EP3000-G2 User Manual

Page 34

Advertising
background image

34

8.2 Technical malfunctions

These malfunctions may occur without an error message being displayed.
*If there are any questions, please contact the Ivoclar Vivadent After Sales Service.

8.3 Repair

Repairs may only be carried out by a certified Ivoclar

Vivadent Service Center. Please refer to the addresses on
the last page of these Operating Instructions.

If repairs during the warranty period are not carried out by a certified
Ivoclar Vivadent Service Center, the warranty will be voided
immediately. Please also refer to the corresponding warranty
regulations.

8.4 Load factory settings

If you want to reset the furnace to its original settings, choose
Settings – Factory Settings (see Chapter 5.5.1). All programs (both
Ivoclar Vivadent and individual programs) and furnace settings will
be reset to the factory settings.

Fehler

Kontrollfrage

Massnahme

Vacuum is not released or only very slowly

Is the vacuum released within approximately 30
seconds?

Wait until the vacuum is released, remove object.
Switch the furnace on and off again. *

Indication on display incomplete

Activate the display test program

*

Writing in the display is very hard to read

Is the contrast properly set?

Adjust contrast

Display not illuminated

Is the furnace properly connected according to the
Operating Instructions and switched on?

Correctly connect the furnace and switch it on

Buzzer does not sound

Is the buzzer switched off (Tune = 0)?

Select tune 1–5

Furnace head does not open

Was the furnace head moved manually?

Open the furnace head only by using the corres-
ponding keys. Switch the furnace on and off again.

Has the vacuum already been released?

Is the program still running? Wait until the program
is complete. Switch furnace off and on again. *

Vacuum pump does not start working

Is the vacuum pump fuse defective?

Check fuse and replace if necessary.

Was the maximum power consumption exceeded?

Use only the vacuum pump recommended by
Ivoclar Vivaden t

Is the vacuum pump plug correctly connected?

Correctly connect the vacuum pump to the furnace
base.

Final vacuum is not reached

Is the vacuum hose OK?

Check vacuum hose and hose connection

Is the pump output OK?

Start the vacuum test program

Humidity/condensation in the vacuum hose?

Start dehumidification program

Incorrect or illogical temperature indication

Is the thermocouple bent or fractured?

*

Is the thermocouple correctly connected?

Correctly connect thermocouple

Is the thermocouple plug defective?

*

Hairline cracks in the heating muffle

Are the cracks very small and insignificant (hairline
cracks)?

Small cracks in the muffle are normal and do not
negatively influence the function of the furnace

Are the cracks large or have parts of the heating
muffle broken off?

*

Cracks in the insulation

Are the cracks very small and insignificant (hairline
cracks)?

Small cracks in the insulation do not negatively
influence the furnace

Are the cracks large or have parts of the insulation
broken off?

*

Cracks in the quartz glass / heating element

Are there cracks in the quartz glass or is the quartz
glass sheathing the heating wire broken?

Switch off the furnace *

Advertising